Common faults and treatment of anaerobic incubator of Xinzhuang Instrument
1. Fault Phenomenon: No Power
a. Cause of the Failure: The power plug is not properly inserted or there is a short circuit. – Troubleshooting: Ensure the plug is fully inserted into the socket. If the problem persists, check for any electrical faults or short circuits in the wiring.
b. Cause of the Failure: The fuse is blown. – Troubleshooting: Replace the fuse with an identical type to restore power supply.
c. Cause of the Failure: The power switch is faulty. – Troubleshooting: Replace the power switch with a new one to ensure proper operation.
2. Fault Phenomenon: Incubator Does Not Open
a. Cause of the Failure: The set temperature is too low. – Troubleshooting: Adjust the set temperature to the required level for proper functioning.
b. Cause of the Failure: The temperature probe is damaged. – Troubleshooting: Replace the faulty temperature probe with a new one to ensure accurate readings.
3. Fault Phenomenon: Air Leakage in the Operation and Sampling Chambers
a. Cause of the Failure: Poor sealing. – Troubleshooting: Use soapy water to locate the leak, then reseal the area with 704 sealant for a tight fit.
b. Cause of the Failure: The rubber band has aged over time. – Troubleshooting: Replace the old rubber band with a new one to maintain proper sealing.
4. Fault Phenomenon: Gas Path Blocked
a. Cause of the Failure: A faulty valve. – Troubleshooting: Replace the defective valve to ensure smooth gas flow.
b. Cause of the Failure: Leaks or disconnection in the leather tube. – Troubleshooting: Replace the damaged leather tube with a special one designed for gas systems.
c. Cause of the Failure: The gas cylinder is empty. – Troubleshooting: Refill or replace the gas cylinder with a new one to resume normal operation.
5. Fault Phenomenon: Light Is Dim or Not Working
a. Cause of the Failure: The starter or lamp is faulty. – Troubleshooting: Replace the defective starter or lamp to restore lighting function.
b. Cause of the Failure: The rectifier is burned out. – Troubleshooting: Replace the damaged rectifier to ensure stable power supply to the lighting system.