Common faults and treatment of anaerobic incubator of Xinzhuang Instrument

1. Fault Phenomenon: No Power

a. Cause of the Failure: The power plug is not properly inserted or there is a short circuit. – Troubleshooting: Ensure the plug is fully inserted into the socket. If the problem persists, check for any electrical faults or short circuits in the wiring.

b. Cause of the Failure: The fuse is blown. – Troubleshooting: Replace the fuse with an identical type to restore power supply.

c. Cause of the Failure: The power switch is faulty. – Troubleshooting: Replace the power switch with a new one to ensure proper operation.

2. Fault Phenomenon: Incubator Does Not Open

a. Cause of the Failure: The set temperature is too low. – Troubleshooting: Adjust the set temperature to the required level for proper functioning.

b. Cause of the Failure: The temperature probe is damaged. – Troubleshooting: Replace the faulty temperature probe with a new one to ensure accurate readings.

3. Fault Phenomenon: Air Leakage in the Operation and Sampling Chambers

a. Cause of the Failure: Poor sealing. – Troubleshooting: Use soapy water to locate the leak, then reseal the area with 704 sealant for a tight fit.

b. Cause of the Failure: The rubber band has aged over time. – Troubleshooting: Replace the old rubber band with a new one to maintain proper sealing.

4. Fault Phenomenon: Gas Path Blocked

a. Cause of the Failure: A faulty valve. – Troubleshooting: Replace the defective valve to ensure smooth gas flow.

b. Cause of the Failure: Leaks or disconnection in the leather tube. – Troubleshooting: Replace the damaged leather tube with a special one designed for gas systems.

c. Cause of the Failure: The gas cylinder is empty. – Troubleshooting: Refill or replace the gas cylinder with a new one to resume normal operation.

5. Fault Phenomenon: Light Is Dim or Not Working

a. Cause of the Failure: The starter or lamp is faulty. – Troubleshooting: Replace the defective starter or lamp to restore lighting function.

b. Cause of the Failure: The rectifier is burned out. – Troubleshooting: Replace the damaged rectifier to ensure stable power supply to the lighting system.

Common Faults and Treatment of Anaerobic Incubator from Xinzhuang Instrument This guide provides a detailed overview of common issues that may occur with anaerobic incubators and offers practical troubleshooting solutions. Whether it's a power issue, temperature control problem, or gas system malfunction, understanding these potential faults can help users quickly identify and resolve them. Regular maintenance and timely replacement of worn parts are essential to keep the incubator running smoothly and efficiently. Always follow safety guidelines when handling electrical components or gas systems.