Common faults and treatment of anaerobic incubator of Xinzhuang Instrument

1. Fault Phenomenon: No Power

a. Cause of the Failure: The power plug is not properly inserted or there is a short circuit. - Troubleshooting: Ensure the plug is fully inserted into the socket. If the issue persists, check for any electrical shorts and repair as necessary.

b. Cause of the Fault: The fuse is blown. - Troubleshooting: Replace the fuse with a new one of the same type and rating.

c. Cause of the Failure: The power switch is faulty. - Troubleshooting: Replace the power switch with a new one to restore functionality.

2. Fault Phenomenon: Incubator Won't Open

a. Cause of the Fault: The set temperature is too low. - Troubleshooting: Adjust the set temperature to a suitable level for your needs.

b. Cause of the Failure: The temperature probe is damaged. - Troubleshooting: Replace the faulty temperature probe with a new one to ensure accurate readings.

3. Fault Phenomenon: Air Leaks in the Operating and Sampling Chambers

a. Cause of the Failure: Poor sealing. - Troubleshooting: Use soapy water to locate the leak, then reseal the area using 704 sealant for a tight fit.

b. Cause of the Failure: Aging rubber band. - Troubleshooting: Replace the old or worn-out rubber band with a new one to prevent air leakage.

4. Fault Phenomenon: Gas Path Blocked

a. Cause of the Failure: Faulty valve. - Troubleshooting: Replace the defective valve to restore proper gas flow.

b. Cause of the Failure: Leaking or disconnected rubber tube. - Troubleshooting: Replace the damaged or loose rubber tube with a special leather tube designed for this purpose.

c. Cause of the Failure: Gas cylinder is empty. - Troubleshooting: Refill or replace the gas cylinder with a full one to resume operation.

5. Fault Phenomenon: Light Is Not Bright

a. Cause of the Failure: The starter or lamp is not working. - Troubleshooting: Replace the faulty starter or the light bulb to restore illumination.

b. Cause of the Failure: The rectifier is burned out. - Troubleshooting: Replace the damaged rectifier to ensure proper power supply to the lighting system.

Common faults and treatment of anaerobic incubator of Xinzhuang Instrument

This guide provides practical troubleshooting steps for common issues that may occur with anaerobic incubators. Whether it's a power problem, temperature control issue, air leakage, gas flow blockage, or lighting failure, understanding the possible causes and their solutions can help maintain the efficiency and reliability of the equipment. Regular maintenance and timely replacement of worn parts are essential to avoid unexpected downtime and ensure safe operation.